We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.

Professional Medical Equipment Supplier - Yuesen Med Professional Medical Equipment Supplier - Yuesen Med

Hot Products

YSX500D 50kW DR system set up and put into service in Cambodia.

YSENMED YSX500D 50kW digital x-ray system has been successfully set up and put into service in a hospital in Cambodia.

YSX056-PE serving as a vehicle-mounted x-ray in the Philippines

YSX056-PE 5.6kW portable x-ray unit has been adapted to fit on a truck, to provide mobile x-ray examination service for remote communities in the Philippines.

X Ray Machine  To  Zimbabwe

x ray machine, 50KW x ray machine

Microscope To Malawi

Achromatic objectives: 4X、10X、40X(S), 100X(S、Oil) Wide field eyepiece: WF10X(WF16X for option) Eyepiece head: Sliding binocular head inclined at 45° Stage: Double layer mechanical stage size 140X140mm, moving range 75X45mm Focusing: Coaxial coarse and

GE Mobile X-ray machine fault

Views : 1565
Author : Monica Liang
Update time : 2015-12-21 08:00:00

The machine over-current protection contactor triped and system was automatic power off when open the machine and pressed positioning lamp or exposure preparation in use.

Failure analysis and maintenance:

AMX-4+ system is consist of power box, charging control, CPU board, high voltage generator, motor control and remote control receiving. Every model is controled and detected by CPU. Then check attached machine circuit diagram and consider that positioning lamp(AC 24V, 150W) and exposure preparation(Rotating anode start ) are provided power by T1 transformer, it is controled by 46-232786G2.

The next step you could take down the board and measure, you would find that F93, which had insurance tube, is burned off and effective tube was breakdown. After replacing the same model tube, connect power and open the machine, fault remains. Then you would doubt CPU control singal abnormal and increase the load spurious. It leads to excessive current and power tube punctured. The following you could use digital multimeter to measure and 60Hz control singal is normal.

Replace the tube again and then connect power, the two test points are 5V and 6V. This two points are not balance. After power off, take down promote tube Q61, there is leakage phenomenon. After replacing the same model tube, connect power and measure two points are 6.0V. After opening the machine and self-inspection, fault is solved and x-ray machine can work normally.

Related News
Read More >>
How to Choose the Right X-Ray Equipment for Your Facility How to Choose the Right X-Ray Equipment for Your Facility
Nov .22.2025
Choosing the right X-ray equipment for your facility can feel like navigating a maze. With so many options available, it's easy to get overwhelmed. In this guide, we'll break down everything you need to know to make an informed decision.
How Many Types of Digital X-Ray Machines? How Many Types of Digital X-Ray Machines?
Nov .15.2025
In this article, we'll explore the various types of digital X-ray machines, their uses, advantages, and how they differ from traditional X-ray systems.
What Are the Advantages of Digital Radiography Compared to Traditional Radiography? What Are the Advantages of Digital Radiography Compared to Traditional Radiography?
Nov .14.2025
What exactly are the advantages of digital radiography compared to its traditional counterpart? Let's dive into this topic and explore the benefits that digital technology brings to the table.
What is the Digital Radiography System? What is the Digital Radiography System?
Nov .09.2025
In this article, we'll dive deep into the world of digital radiography, exploring its components, advantages, and applications.